Operation and handling
Description of the LED combination LED (C) blue, LED (D) red "Reagent message / BoB message".
LED display
(C) and (D)
Description
Blue-glowing
The indicator lights up permanently, signalling that the reagent
supply is less than approx. 30 %. The remaining amount of reagent
allows operation for another 72 hours in the set analysis interval.
Prepare a new reagent bottle.
LED (C) Blue flashing
LED (D) Red-flashing
The BOB alarm is active. The display signals a reagent level that may
no longer be sufficient for unattended operation in the set analysis
interval (BoB) over 72 hours.
Relay 2 "Device fault / reagent message BoB " has dropped out.
To acknowledge the BoB alarm, press the "Reset" button.
LED (C) Blue flashing
LED (D) Red-glowing
The indicator signals a reagent level that may no longer be
sufficient for unattended operation for 72 hours.
The BoB message has already been acknowledged with the
"Reset" button.
Relay 2 "Device fault / reagent message BoB " is energised.
Description of the red LED (D) "Device malfunction
LED display (D)
Description
Flashing red
The display signals a unit malfunction:
•
Faulty zero sample or faulty measurement
•
All other displays are switched off.
To acknowledge the unit fault, press the "Reset" button or switch
the unit off and on again.
Relay 2 "Device fault / reagent message BoB " has dropped out.