Operation and handling
If the input contact is permanently bridged while the programme switch SW3 is ON,
analyses are carried out continuously. The analyses are carried out consecutively.
The set internal analysis interval is deactivated.
Symbols used:
INPUT
Input contact
State
Description
Closed
The input contact is permanently closed
Open
The input contact is permanently open
Pulse
Only one analysis is carried out
Setting for the use of an external control depending on SW1 and SW2 switch position
External analysis start,
input contact activated (SW3 ON)
SW programme
switch position
INPUT
Input contact
Function
For pulsed
operation
Analyses are carried out continuously without pauses
Paused: No analyses are carried out because the input
contact is open.
Only one analysis is carried out
Analyses are carried out continuously without pauses
Analyses are carried out at
10 min intervals
Only one analysis is carried out
Analyses are carried out continuously without pauses
Analyses are carried out at
20 min intervals
Only one analysis is carried out
Analyses are carried out continuously without pauses
Analyses are carried out at
30 min intervals
Only one analysis is carried out