Electrical installation
Wiring Example 6: LIMESBase
Attention
•
Relay output 1
picked up
when limit value is exceeded
(COM, NO)
•
Relay output 2 in case of unit malfunctions or reagent message BoB
dropped out
(COM,
NC)
The input contact function is not active on the LIMESBase. Controls or switches connected to the
input contact do not cause analyses to start or stop. To enable the input contact function, you need
an upgrade kit, which you can purchase from us. For
more information, see page 8, 21 and 68.
Drawn position for relay 1 and relay 2: Unit de-energised - Relay de-energised
LIMES
B
as
e
Jun
ct
ion
box
(o
pt
ion
al)
Illustration 18: Circuit diagram 6 - LIMESBase
Relay 2 is always energized
during normal operation:
Connection COM to NO
(wire breakage safety)