Performance-Test
Use Suspension for Performance-Test?:
Carry out the Particle Counting Accuracy stipulated by the USP.
Select this to use the Suspension.
Use Blank Bottle?:
Select this to measure and subtract the Blank Bottle counts (which is
part of the Performance-Test set of USP).
Lower Threshold (µm):
Set the lower threshold.
The allowable setting range is 1.0 to 99.9 µm (0.1 µm step).
Upper Threshold (µm):
Set the upper threshold.
The allowable setting range is 1.1 to 100.0 µm (0.1 µm step).
Minimum Concentration (/mL):
Set the minimum number concentration of the test suspension. The allow-
able setting range is 100 to 9999 particles / mL (1 particle / mL step).
Maximum Concentration (/mL):
Set the maximum number concentration of the test suspension. The allow-
able setting range is 101 to 10000 particles / mL (1 particle / mL step).
Minimum Ratio:
Set the minimum ratio according to USP.
The allowable setting range is 1.00 to 4.99 (0.01 step).
Maximum Ratio:
Set the maximum ratio according to USP.
The allowable setting range is 1.01 to 5.00 (0.01 step).
ID of the Suspension:
Input a Suspension ID with less than 18 letters / fi gures.
Suspension