72
Performance-Test
Calibration Data
Calibration Data List:
The registered Calibration Data name list will be displayed.
Select the appropriate Calibration data to carry out a Performance-Test.
Note
Please note that an incorrect data selection will not enable you to
carry out a correct Performance-Test.
The following data are registered at the dispatch of this unit.
KL-04_Basic:
Calibrated according to a method in line with RION’s specifi ed
method.
JPxx:
Calibrated according to a method in accordance with the
Japanese Pharmacopeia (JP).
(Registered only for those who have ordered Option(s))
USPxx:
Calibrated using a method in accordance with the United States
Pharmacopeia (USP).
(Registered only for those who have ordered Option(s))
EPxx:
Calibrated using a method in accordance with European Phar-
macopeia (EP).
(Registered only for those who have ordered Option(s))
* xx indicates each pharmacopeia’s revision numbers (JP and USP) or ver-
sion numbers (EP).
EP has no regulations about Calibration and Performance-Test, so will be
applied JP or USP by customer specifi cation.
“View” button:
This enables checking of the selected Calibration Data.
Select a required calibration data name from the list then click this button.
The “KL-04 - Calibration” window appears.