22
Preparation
How to use
1. Ensure that the volume dial on the controller is turned OFF.
Completely turn the volume dial to the left until it stops as per the
diagram on the right.
2. Connect the power cord to the AC outlet.
3. Put a stirring bar into the sample container and set the sample
container on the sample stand.
Caution
The sample container must be set on the sample stand
after ensuring the volume dial on the controller is OFF.
If the sample container is set when the volume dial is
ON, the stirring bar may jump around, which can cause
splashing of the fl uid and damage to the sample con-
tainer. Also the level of fl uid in the sample container
must be controlled to prevent it from overfl owing when
the stirring bar is revolving at high speed. The standard
level of sample fl uid is 80% of the displayed capacity
of the sample container.
The surface material of the provided stirring bar is glass.
Some sample fl uids can cause corrosion, dissolving and
other chemical reactions to the stirring bar. Ingredients
of sample fl uids must be checked before they are put
in the sample container.
If the stirring bar is not cleaned adequately contamination
of sample fl uids and chemical reactions can occur. The
cleanness of the stirring bar must be ensured before
putting into the sample container.