Performance-Test
6. Set a sample (in general purifi ed water) for cleaning purposes, then click the “Flush”
button on the display or press the “FLUSH” button on the front of the unit.
Important
When operating by pressing the “FLUSH” and “START”
buttons on the unit, ensure that the “KL-04 - Performance
Test” window is set in an active window in which the required
Performance-Test Parameter Name is displayed. If it is not
in an active window, the operations may be different from
the specifi ed Parameter.
7. Carry out the Performance-Test. Prepare the sample(s) then click a button for the
required test item (e.g. 10.350 µm) shown on the list.
After the test is completed, the test results will be displayed to the right of the
items.
Note
The test result will be displayed in black if accepted by the system,
and in red if rejected.
8. If more test items are required, repeat the sequence from No. 6 to 7 for the number
of test items only.
After all test items are completed, click the “Register” button to register the test
result. The Registration is not available for rejected test results.
9. Click the “Quit” button upon completion of the test.
Note
The Performance-Test series should be fully completed within
72 hours. If this time limit is exceeded, the test result becomes
void.