Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
915
User Manual 1178.8520.02 ─ 01
ON | 1
Switches the function on
*RST:
0
Example:
SWE:EGAT ON
Activate gating
SWE:EGAT:CONT:STAT ON
Activate continuous gating
Manual operation:
See
[SENSe:]SWEep:EGATe:HOLDoff
<DelayTime>
This command defines the delay time between the gate signal and the continuation of
the measurement.
Note:
If you perform gated measurements in combination with the IF Power trigger, the
R&S
FSV/A ignores the holding time for frequency sweep, FFT sweep, zero span and
I/Q mode measurements.
Parameters:
<DelayTime>
Range:
0 s to 30 s
*RST:
0 s
Default unit: S
Example:
SWE:EGAT:HOLD 100us
Manual operation:
See
[SENSe:]SWEep:EGATe:LENGth
<GateLength>
This command defines the gate length.
Parameters:
<GateLength>
Range:
125 ns to 30 s
*RST:
400μs
Default unit: S
Example:
SWE:EGAT:LENG 10ms
Manual operation:
See
[SENSe:]SWEep:EGATe:POLarity
<Polarity>
This command selects the polarity of an external gate signal.
The setting applies both to the edge of an edge-triggered signal and the level of a
level-triggered signal.
Parameters:
<Polarity>
POSitive | NEGative
*RST:
POSitive
Example:
SWE:EGAT:POL POS
Setting Basic Measurement Parameters