Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
210
User Manual 1178.8520.02 ─ 01
Retrieving results via remote control
The measurement results of the spectrum emission mask
test
can be retrieved using
the
CALC:LIM:FAIL?
command from a remote computer; see
on page 1055 for a detailed description.
The
power
result for the reference range can be queried using
CALC:MARK:FUNC:POW:RES? CPOW
;
The
peak
power for the reference range can be queried using
CALC:MARK:FUNC:POW:RES? PPOW
, see
CALCulate<n>:MARKer<m>:FUNCtion:
The measured power
trace
can be queried using
TRAC:DATA?
and
TRAC:DATA:X?
,
The measured
peak power list
can be queried using
TRAC:DATA? LIST
, see
7.6.4
SEM Basics
Some background knowledge on basic terms and principles used in SEM measure-
ments is provided here for a better understanding of the required configuration set-
tings.
................................................................................. 210
Limit Lines in SEM Measurements
........................................................................212
......................................................................................214
Multi-Standard Radio (MSR) SEM Measurements
............................................... 216
SEM with Multiple Sub Blocks ("Multi-SEM")
........................................................ 216
7.6.4.1
Ranges and Range Settings
In the Spectrum Emission Mask measurements, a range defines a segment for which
you can define the following parameters separately:
●
Start and stop frequency
●
RBW
●
VBW
●
"Sweep Time"
●
"Sweep Points"
●
Reference level
●
Attenuator settings
●
Preamplifier settings
●
Transducer settings
●
Limit values
Via the sweep list, you define the ranges and their settings. For details on settings,
refer to
Chapter 7.6.5.1, " Sweep List "
Spectrum Emission Mask (SEM) Measurement