Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1109
User Manual 1178.8520.02 ─ 01
ALL
The frequency response correction settings are applied to all
active measurement channels.
Example:
INST:SEL 'MyIQSpectrum'
SENS:CORR:FRES:USER:SCOP CHAN
The filter is applied only to the channel named 'MyIQSpectrum'.
Manual operation:
See
[SENSe:]CORRection:FRESponse:BASeband:USER:SLISt<sli>:CATalog?
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:SLISt<sli>:CATalog?
[SENSe:]CORRection:FRESponse:USER:SLISt<sli>:CATalog?
Returns a list of currently configured touchstone format files.
Suffix:
<sli>
.
1..n
Index in Touchstone file list
Use
[SENSe:]CORRection:FRESponse:USER:
to determine the maximum index number.
Return values:
<FileList>
path and file name of all currently configured touchstone format
files.
Example:
SENS:CORR:FRES:USER:SLIS:CAT?
C:\R_S\INSTR\USER\MyS7p.s7p,C:
\R_S\INSTR\USER\MyS2p.s2p
Usage:
Query only
[SENSe:]CORRection:FRESponse:BASeband:USER:SLISt<sli>:CLEar
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:SLISt<sli>:CLEar
[SENSe:]CORRection:FRESponse:USER:SLISt<sli>:CLEar
Removes
all
touchstone files from the current configuration for the selected or all input
types.
Suffix:
<sli>
.
irrelevant
Example:
SENS:CORR:FRES:USER:SLIS:CLE
Usage:
Event
Manual operation:
See
[SENSe:]CORRection:FRESponse:USER:SLISt<sli>:DATA:FREQuency<spi>?
[SENSe:]CORRection:FRESponse:USER:SLISt<sli>:DATA:MAGNitude<spi>?
[SENSe:]CORRection:FRESponse:USER:SLISt<sli>:DATA:PHASe<spi>?
Queries the trace values for the specified
.snp
file and ports.
Configuring the R&S