Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
220
User Manual 1178.8520.02 ─ 01
Figure 7-39: Summarized limit line for multiple sub blocks
7.6.5
SEM Configuration
Access
: "Overview" > "Select Measurement" > "Spectrum Emission Mask"
The SEM measurement is started immediately with the default settings.
Auto measurement
Use the
function to determine the required measurement parameters automat-
ically.
The following settings are determined:
●
Center frequency
●
Span: determined signal bandwidth plus 40
%
●
SEM standard according to estimated signal bandwidth
●
For pulsed signals: gating parameters
See also
Chapter 7.1, "Auto Measurements"
The remote commands required to perform these tasks are described in
ter 14.5.6, "Measuring the Spectrum Emission Mask"
Global span settings
The span of the signal to be monitored is configured in the general span settings (see
Chapter 8.3.2, "Frequency and Span Settings"
on page 360). Only ranges within this
global span are considered for the SEM measurement.
Spectrum Emission Mask (SEM) Measurement