Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
934
User Manual 1178.8520.02 ─ 01
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet
<Offset>
This command defines a level offset for the external generator level. Thus, for exam-
ple, attenuators or amplifiers at the output of the external generator can be taken into
account for the setting.
Suffix:
<si>
.
irrelevant
Parameters:
<Offset>
Range:
-200 dB to +200 dB
*RST:
0dB
Default unit: DB
Example:
//Define a level offset on the external generator
SOUR:POW:OFFS -10dB
Manual operation:
See
Interface Configuration
The following commands are required to configure the interface for the connection to
the external generator.
SOURce<si>:EXTernal<gen>:ROSCillator[:SOURce]
.........................................................934
SYSTem:COMMunicate:GPIB:RDEVice:GENerator<gen>:ADDRess
...................................935
SYSTem:COMMunicate:RDEVice:GENerator<gen>:INTerface
............................................935
SYSTem:COMMunicate:RDEVice:GENerator<gen>:LINK
.................................................. 935
SYSTem:COMMunicate:RDEVice:GENerator<gen>:TYPE
................................................. 936
SYSTem:COMMunicate:TCPip:RDEVice:GENerator<gen>:ADDRess
................................. 936
SOURce<si>:EXTernal<gen>:ROSCillator[:SOURce]
<Source>
This command controls selection of the reference oscillator for the external generator.
If the external reference oscillator is selected, the reference signal must be connected
to the rear panel of the instrument.
Suffix:
<si>
.
irrelevant
<gen>
irrelevant
Parameters:
<Source>
INTernal
Uses the internal reference.
EXTernal
Uses the external reference; if none is available, an error flag is
displayed in the status bar.
*RST:
INT
Example:
//Select an external reference oscillator
SOUR:EXT:ROSC EXT
Manual operation:
See
Setting Basic Measurement Parameters