General Instrument Setup
R&S
®
FSVA3000/ R&S
®
FSV3000
628
User Manual 1178.8520.02 ─ 01
12.9.3
Calibration Signal Display
Access
: [Setup] > "Service" > "Calibration Signal"
Alternatively to the RF input signal from the front panel connector you can use the
instrument's calibration signal as the input signal, for example to perform service func-
tions on.
.......................................................................................... 628
.........................................................................................629
NONE
Uses the current RF signal at the input, i.e. no calibration signal (default).
Remote command:
DIAGnostic:SERVice:INPut[:SELect]
Calibration Frequency RF
Uses the internal calibration signal as the RF input signal.
Remote command:
DIAGnostic:SERVice:INPut[:SELect]
DIAGnostic:SERVice:INPut:PULSed:CFRequency
Spectrum ← Calibration Frequency RF
Defines whether a broadband or narrowband calibration signal is sent to the RF input.
"Narrowband"
Used to calibrate the absolute level of the frontend at 64 MHz.
Service Functions