Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
407
User Manual 1178.8520.02 ─ 01
The gate length can only be set in the edge-triggered gate mode. In the level-triggered
mode the gate length depends on the level of the gate signal.
The gate length in relation to the sweep is indicated by a line labeled "GL" .
For more information see
Chapter 8.6.2.1, "Gated Measurements"
Remote command:
8.6.2.3
Continuous Gate Settings
Access
: "Overview" > "Trigger" > "Trigger / Gate Config." > "Cont. Gate" tab
Continuous gating allows you to perform a continuous gated sweep after a single
external, IF power or power sensor trigger is received.
For details see
Continuous Gate
Activates or deactivates continuous gating.
This setting is only available if
Remote command:
[SENSe:]SWEep:EGATe:CONTinuous[:STATe]
Gate Period Length
Defines the length in seconds of a single gate period in continuous gating. The length
is determined from the beginning of one gate measurement to the beginning of the
next one.
Remote command:
[SENSe:]SWEep:EGATe:CONTinuous:PLENgth
Gate Period Count
Defines the number of gate periods to be measured after a single trigger event in con-
tinuous gating.
Trigger and Gate Configuration