Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
130
Operating Manual 1176.7561.02 ─ 06
Example:
[SENSe:]ADEMod:FM[:TDOMain][:TYPE]
"Creates average, max hold and min hold values simultaneously
DEM:FM WRIT,OFF,OFF
Only creates the current measurement values
ADEM:FM OFF,OFF,OFF
Switches analog demodulator off
Mode:
ADEMOD
[SENSe:]ADEMod:FM[:TDOMain]:RESult
<Type>
This command reads the result data obtained by analog demodulation for the specified
result type. The data format of the output data block is defined by the FORMat com-
mand.
Return values:
<Type>
<result type>: WRITe, AVERage, MAXHold, MINHold; for details
see
Chapter 2.3.6.1, "Trace Mode Result Types"
The result type indicated must be one of those configured by
[SENSe:]ADEMod:FM[:TDOMain][:TYPE]
Otherwise a query error will be generated.
Return values
ASCII Format (FORMat ASCII) or Binary Format (FORMat
REAL,32); for details see
Returned Values: ASCII Format and Binary Format"
on page 118 .
Default unit: Hz
Example:
ADEM:SET 8MHz,32000,EXT,POS,-500,30
, see
Sets up demodulator parameters
ADEM:FM AVER,MAXH,MINH
, see
Selects FM results to be measured
ADEM:AM WRIT,OFF,OFF
Selects AM results to be measured
ADEM ON
, see
Switches on demodulator
INIT; WAI
Starts measurement and waits for sync
FORM ASC
, see
Selects output format
ADEM:FM:RES? AVER
Reads FM average results
ADEM:FM:RES? MAXH
Reads FM max hold results
ADEM:FM:RES? MINH
Reads FM min hold results
ADEM:AM:RES? WRIT
Reads current AM results
Remote Commands of the Analog Demodulation (R&S
FSV–K7)