Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
134
Operating Manual 1176.7561.02 ─ 06
[SENSe:]ADEMod:PM[:TDOMain][:TYPE]
<Type>
This command selects the result types of the PM-demodulated signal to be created
simultaneously.
Parameters:
<Type>
*RST:
OFF,OFF,OFF
<result type 1|2|3|4|5|6>: WRITe, AVERage, MAXHold, MINHold,
VIEW; for details see
Chapter 2.3.6.1, "Trace Mode Result
Example:
ADEM:PM AVER,MAXH,MINH
, see
Determines average, maximum and minimum value simultane-
ously
ADEM:PM WRIT,OFF,OFF
Determines only current measurement results
ADEM:PM OFF,OFF,OFF
Switches the PM demodulator off.
Mode:
ADEMOD
[SENSe:]ADEMod:PM[:TDOMain]:RESult
<Type>
This command reads the result data of the PM demodulation for the specified result
type. The data format of the output data is determined with the FORMat command.
Return values:
<Type>
<result type>: WRITe, AVERage, MAXHold, MINHold; for details
see
Chapter 2.3.6.1, "Trace Mode Result Types"
The specified result type must be one of those configured with
the
[SENSe:]ADEMod:PM[:TDOMain][:TYPE]
Otherwise a query error will be generated.
Return values
ASCII Format (FORMat ASCII) or Binary Format (FORMat
REAL,32); for details see
Returned Values: ASCII Format and Binary Format"
on page 118 .
Default unit: dB (logarithmic display) or RAD or DEG (linear dis-
play)
Remote Commands of the Analog Demodulation (R&S
FSV–K7)