Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
100
Operating Manual 1176.7561.02 ─ 06
Example:
CALC:MARK:FUNC:FPE:SORT Y
Sets the sort mode to decreasing y values
CALC:MARK:FUNC:FPE 3
Searches the 3 highest maxima for trace 1
CALC:MARK:FUNC:FPE:COUN?
Queries the number of maxima found
CALC:MARK:FUNC:FPE:X?
Queries the frequencies (span <> 0) or. time (span = 0) of the
maxima found
107.5E6,153.8E6,187.9E6
frequencies in increasing order
2.05E-3,2.37E-3, 3.71e-3
times in increasing order
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:Y?
This command queries the position of the peaks on the y-axis.
The order depends on the sort order that has been set with
MARKer<m>:FUNCtion:FPEaks:SORT
The number of peaks on the number that has been set with
MARKer<m>:FUNCtion:FPEaks:COUNt?
.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Return values:
<PeakPosition>
Position of the peaks on the y-axis. The unit depends on the
measurement.
Example:
CALC:MARK:FUNC:FPE:SORT Y
Sets the sort mode to decreasing y values
CALC:MARK:FUNC:FPE 3
Searches the 3 highest maxima for trace 1
CALC:MARK:FUNC:FPE:COUN?
Queries the number of maxima found
CALC:MARK:FUNC:FPE:Y?
Queries the levels of the maxima found
Usage:
Query only
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown
<Distance>
This command defines the distance of the n dB down markers to the reference marker.
Remote Commands of the Analog Demodulation (R&S
FSV–K7)