Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
146
Operating Manual 1176.7561.02 ─ 06
Parameters:
<Threshold>
Range:
0 to 200
*RST:
+1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:LOW 2
Example:
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level falls below 18
dBm.
Manual operation:
See
[SENSe:]ADJust:CONFiguration:HYSTeresis:UPPer
<Threshold>
This command defines an upper threshold the signal must exceed before the reference
level is automatically adjusted when the "Auto Level" function is performed.
For more information see
Parameters:
<Threshold>
Range:
0 to 200
*RST:
+1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:UPP 2
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level rises above 22
dBm.
Manual operation:
See
[SENSe:]ADJust:CONFigure:LEVel:DURation
<Duration>
This command defines the duration of the level measurement used to determine the
optimal reference level automatically (for
SENS:ADJ:LEV ON
).
Parameters:
<Duration>
<numeric value> in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s
Example:
ADJ:CONF:LEV:DUR:5
Manual operation:
See
[SENSe:]ADJust:FREQuency
This command defines the center frequency and the reference level automatically by
determining the highest level in the frequency span.
Example:
ADJ:FREQ
Manual operation:
See
Remote Commands of the Analog Demodulation (R&S
FSV–K7)