Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
182
Operating Manual 1176.7561.02 ─ 06
This command is not available when using R&S Digital I/Q Interface (R&S
FSV-B17).
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
INP:GAIN:STAT ON
Turns the preamplifier on.
Manual operation:
See
INPut:IMPedance
<Impedance>
This command selects the nominal input impedance.
75
Ω
should be selected if the 50
Ω
input impedance is transformed to a higher impe-
dance using a 75
Ω
adapter of the RAZ type (= 25
Ω
in series to the input impedance
of the instrument). The correction value in this case is 1.76 dB = 10 log (75
Ω
/50
Ω
).
This function is not available if the R&S Digital I/Q Interface (R&S
FSV-B17) is active.
Parameters:
<Impedance>
50 | 75
*RST:
50 Ω
Example:
INP:IMP 75
Manual operation:
See
INPut:SELect
<Source>
This command selects the signal source for measurements.
Parameters:
<Source>
RF
Radio Frequency ("RF INPUT" connector)
DIQ
Digital IQ (only available with R&S Digital I/Q Interface, option
R&S
FSV-B17)
*RST:
RF
Example:
INP:SEL RF
Mode:
A, IQ, NF, TDS, VSA, CDMA, EVDO, WCDMA, ADEMOD, GSM,
OFDM, OFDMA/WiBro, WLAN
Manual operation:
See
Remote Commands of the Analog Demodulation (R&S
FSV–K7)