Operating Manual 1176.7561.02 ─ 06
CALCulate<n>:DELTamarker<m>:Y?
This command queries the measured value of a delta marker. The corresponding delta
marker is activated, if necessary. The output is always a relative value referred to
marker 1 or to the reference position (reference fixed active).
To obtain a correct query result, a complete sweep with synchronization to the sweep
end must be performed between the activation of the delta marker and the query of the
y value. This is only possible in single sweep mode.
Depending on the unit defined with
CALC:UNIT:POW
or on the activated measuring
functions, the query result is output in the units below:
Table 2-6: Analog demodulation measurements
Parameter, measuring function or result display
Output unit
AM result display (R&S
FSV–K7)
% (lin)
dB (log)
FM result display (R&S
FSV–K7)
Hz (lin)
dB (log)
PM result display (R&S
FSV–K7)
rad | deg (lin)
dB (log)
RF result display (R&S
FSV–K7)
dB (Range Log or Range Linear %)
% (Range Linear %)
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Example:
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
CALC:DELT2 ON
Switches on delta marker 2.
CALC:DELT2:Y?
Outputs measurement value of delta marker 2.
Usage:
Query only
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
2.3.10.2
INPut subsystem
............................................................................................... 177
.......................................................................................... 179
Remote Commands of the Analog Demodulation (R&S
FSV–K7)