Mode:
SFM
Manual operation:
See
TRIGger<n>[:SEQuence]:SOURce
<Source>
This command selects the trigger source for the start of a sweep.
Suffix:
<n>
.
1...4
irrelevant
Parameters:
<Source>
IMMediate | EXTern | IFPower | AF | FM | AM | AMRelative |
PM | TIME | SLEFt | SRIGht | SMPX | SMONo | SSTereo |
SRDS | SPILot
For details on trigger sources refer to the
on page 216 softkey.
For triggering with AF, AM, AMRelative, FM, and PM trigger
sources to be successful, the measurement time must cover at
least 5 periods of the audio signal.
IMMediate
Free Run (no trigger)
EXTern
External trigger
IFPower
Triggering via signals which are outside the measurement chan-
nel
AF
Audio frequency trigger
FM
Triggering via FM frequency level
AM
Triggering via RF power signal
AMRelative
Triggering via AM signal
PM
Triggering via PM frequency level
TIME
Triggering according to repetition interval
SLEFt
Triggering via left stereo signal
SRIGht
Triggering via right stereo signal
SMPX
Triggering via MPX stereo signal
Remote Commands of the FM Stereo Option (R&S