Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
14
Operating Manual 1176.7561.02 ─ 06
Option K7S
The K7S option always uses the demodulation bandwidth 400 kHz.
Table 2-1: Sample Rate, Measurement Time and Trigger Offset using a flat demodulation filter
Demod. band-
width
Sample rate
Measurement time
Trigger offset
Min.
Max.
Min.
Max.
160 MHz
1)
200 MHz
5 ns
8 ms
-8 ms
1.0486 s
80 MHz
2)
128 MHz
7.8125 ns
12.5 ms
-12.5 ms
1.6384 s
40 MHz
2)
64 MHz
15.625 ns
25 ms
-25 ms
3.2768 s
28 MHz
64 MHz
15.625 ns
25 ms
-25 ms
3.2768 s
18 MHz
32 MHz
31.25 ns
50 ms
-50 ms
6.5536 s
10 MHz
32 MHz
31.25 ns
50 ms
-50 ms
6.5536 s
8 MHz
16 MHz
62.5 ns
100 ms
-100 ms
13.1072 s
5 MHz
8 MHz
125 ns
200 ms
-200 ms
26.2144 s
3 MHz
4 MHz
250 ns
400 ms
-400 ms
52.4288 s
1.6 MHz
2 MHz
500 ns
800 ms
-800 ms
104.8576 s
800 kHz
1 MHz
1
µ
s
1.6 s
-1.6 s
209.7152 s
400 kHz
500 kHz
2
µ
s
3.2 s
-3.2 s
419.4304 s
200 kHz
250 kHz
4
µ
s
6.4 s
-6.4 s
838.8608 s
100 kHz
125 kHz
8
µ
s
12.8 s
-12.8 s
1677.7216 s
50 kHz
62.5 kHz
16
µ
s
25.6 s
-25.6 s
3355.4432 s
25 kHz
31.25 kHz
32
µ
s
51.2 s
-51.2 s
6710.8864 s
12.5 kHz
15.625 kHz
64
µ
s
102.4 s
-102.4 s
13421.7728 s
6.4 kHz
7.8125 kHz
128
µ
s
204.8 s
-204.8 s
26843.5456 s
3.2 kHz
3.90625 kHz
256
µ
s
409.6 s
-409.6 s
53687.0912 s
1.6 kHz
1.953125 kHz
512
µ
s
819.2 s
-819.2 s
107374.1824 s
800 Hz
976.5625 Hz
1.024 ms
1638.4 s
-1638.4 s
214748.3648 s
400 Hz
488.28125 Hz
2.048 ms
3276.8 s
-3276.8 s
429496.7296 s
200 Hz
244.140625 Hz
4.096 ms
6553.6 s
-6553.6 s
858993.4592 s
100 Hz
122.0703125
Hz
8.192 ms
13107.2 s
-13107.2 s
1717986.918 s
1)
only available with option R&S
FSV-B160
2)
only available with option R&S
FSV-B70/R&S
FSVA-B40
Instrument Functions Analog Demodulation