Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
142
Operating Manual 1176.7561.02 ─ 06
[SENSe:]ADEMod:SPECtrum:SPAN:ZOOM
<FreqRange>
This command sets the frequency range for the RF spectrum result display determined
from analog demodulation data. The frequency range for result display is limited to the
maximum span (
[SENSe:]ADEMod:SPECtrum:SPAN[:MAXimum]
to the measurement bandwidth of analog demodulation (
Parameters:
<FreqRange>
*RST:
5 MHz
Example:
ADEM ON
, see
Switches on the analog demodulator
CALC:FEED 'XTIM:SPEC'
, see
on page 90
Switches on RF spectrum result display.
ADEM:SPEC:SPAN:MAX 5 MHz
, see
Sets the maximum span to 5 MHz
ADEM:SPEC:SPAN:ZOOM 1 MHz
Sets displayed span to 1 MHz
Mode:
ADEMOD
Manual operation:
See
See
[SENSe:]ADEMod:SQUelch[:STATe]
<State>
This command activates the squelch function, i.e. if the signal falls below a defined
threshold (see
data is automatically set to 0.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
DEM:SQU ON
Signals below the level threshold are squelched.
Usage:
SCPI confirmed
Mode:
A, ADEMOD, SFM
Manual operation:
See
[SENSe:]ADEMod:SQUelch:LEVel
<Threshold>
This command defines the level threshold below which the demodulated data is set to
0 if squelching is enabled (see
[SENSe:]ADEMod:SQUelch[:STATe]
Remote Commands of the Analog Demodulation (R&S
FSV–K7)