Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
138
Operating Manual 1176.7561.02 ─ 06
Usage:
SCPI confirmed
Mode:
ADEMOD
Manual operation:
See
[SENSe:]ADEMod:RLENgth?
This command returns the record length set up for the current analog demodulation
measurement.
Example:
ADEM:RLEN?
Returns the current record length.
Usage:
Query only
Mode:
ADEMOD
[SENSe:]ADEMod:SET
<sample rate> | <record length> | <trigger source> | <trigger
slope> | <offset samples> | <# of meas>
This command configures the analog demodulator of the instrument.
Parameters:
<sample rate>
numeric value
The frequency at which measurement values are taken from the
A/D-converter and stored in I/Q memory.
Allowed range: refer to
Sample Rate, Measurement Time and
Note:
for FM stereo measurements (K7S option), the sample
rate is always 500 kHz (as the demodulation bandwidth is per-
manently set to 400 kHz). Thus, this parameter is ignored in this
case.
*RST:
8 MHz
<record length>
Number of samples to be stored in I/Q memory.
Range:
1 to 400001 with AF filter or AF trigger active, 1 to
480001 with both AF filter and AF trigger deactive
*RST:
501)
<trigger source>
Selection of the trigger source to use for the demodulator.
IMMediate | EXTernal | IFPower | RFPower| AF | AM | AMRe-
lative | FM | PM
Note:
After selecting IF Power, the trigger threshold can be set
with the
TRIGger<n>[:SEQuence]:LEVel:IFPower
mand.
*RST:
IMMediate
Remote Commands of the Analog Demodulation (R&S
FSV–K7)