TRACe - Subsystem
R&S FSMR
1313.9723.12
6.1-230
E-1
ASCII Format (FORMat ASCII):
In this case the command returns a comma separated list of the measured voltage values in floating
point format (Comma Separated Values = CSV). The number of values returned is 2 * number of
samples, the first half being the I-values, the second half the Q-values.
Binary Format (FORMat REAL,32):
In this case the command returns binary data (Definite Length Block Data according to IEEE 488.2),
with the lists of I- and Q-data being arranged one after the other in 32 Bit IEEE 754 floating point
data. The scheme of the response string is as follows:
#44096<I-value1><I-value2>...<I-value512><Q-value1><Q-value2>...<Q-value512>
with
#4
digits of the subsequent number of data bytes (4 in the example)
4096
number of subsequent data bytes (
# of DataBytes
, 4096 in the example)
<I-value x>
4-Byte-Floating Point I-value
<Q-value y>
4-Byte-Floating Point Q-value
The number of I- and Q-data can be calculated as follows:
8
#
#
#
DataBytes
of
Data
Q
of
Data
I
of
=
=
The offset of Q-data in the output buffer can be calculated as follows:
s
catorDigit
LengthIndi
DataBytes
of
Offset
Data
Q
+
=
2
)
(#
with LengthIndicatorDigits being the number of digits of the length indicator including the '#'. In the
example above (#44096...) this results in a value of 6 for LengthIndicatorDigits and the offset for the
Q-data will result in 2048 + 6 = 2054.
Characteristics
: *RST value:
--
Note:
Using the command with the *RST values for command
TRAC:IQ:SET the following minimum buffer sizes for
the response data are recommended:
ASCII format: 10 kBytes
Binary format: 2 kBytes
SCPI:
device specific
The command requires that all response data are read out completely before the
instrument accepts further commands.
TRACe<1|2>:IQ:DATA:FORMat
COMPatible
| IQBLock | IQPair
This command sets the data output formatting (using the TRAC:IQ:DATA? command):
COMPatible
512k I data and 512k Q data is alternately transferred
IQBLock
First, all I and then all Q data is transferred
IQPair
I/Q data is transferred in pairs
Example:
"TRAC:IQ:DATA:FORM IQP"
Features:
*RST value:
COMP
SCPI:
device-specific
Содержание FSMR
Страница 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Страница 4: ......
Страница 22: ......
Страница 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Страница 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Страница 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Страница 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Страница 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Страница 66: ......
Страница 86: ......
Страница 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Страница 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Страница 144: ......
Страница 150: ......
Страница 156: ......
Страница 202: ......
Страница 208: ......
Страница 216: ......
Страница 230: ......
Страница 254: ......
Страница 280: ......
Страница 348: ......
Страница 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Страница 390: ......
Страница 408: ......
Страница 428: ......
Страница 446: ......
Страница 536: ......
Страница 799: ......
Страница 921: ......
Страница 935: ......
Страница 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...