R&S FSMR
SENSe:LIST Subsystem
1313.9723.12
6.1-171
E-1
<ref level>:
Reference level
Range of values: +30 dBm to -70 dBm in 10 dB steps
+30 dBm to -75 dBm in 5 dB steps with
El. Attenuator option B25
Range of values: 0 dB to 70 dB
in 10 dB steps
0 dB to 75 dB
in 5 dB steps with
El. Attenuator Option
B25
<el att>:
RF input attenuation of electronic attenuator
Range of values: 0 dB to 30 dB
in 10 dB steps
OFF
electronic attenuator
not in signal path
If option B25 is missing, OFF is to be used.
<filter type>:
NORMal:
normal resolution filter
CFILter:
channel filter. These are especially steep-edged
filters, which are used for example in Fast ACP
measurement to ensure the band-limiting of a
transmission channel in the time domain.
RRC:
Root Raised Cosine filter. This special filter form
is used to determine the channel power for some
mobile radio standards.
<rbw>:
Resolution bandwidth
Range of values: 10 Hz to 10 MHz in 1, 3, 10 steps and
50 MHz (R&S FSMR43: 10Hz to 10MHz)
for <filter type> = NORMal.
<vbw>:
Video bandwidth
Range of values: 1 Hz to 10 MHz in 1, 2, 3, 5, 10 steps.
(30 MHZ for RBW > 10 MHz, not R&S FSMR)
The value is ignored for
<filter type> =
CFILter or RRC
<meas time>:
Measurement time
Range of values:
1us to 30s
<trigger level>:
Reserved. Must be set to 0.
Returned values:
The query command returns a list of comma-separated values (CSV) which contains the power
measurement results in floating-point format. The unit depends on the setting with
CALC:UNIT.
Command
"SENSe:LIST:POWer?
935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,10dB,NORM,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,20dB,NORM,30kHz,100kHz,434us,0"
thus returns the following list, for example:
-28.3,-30.6,-38.1
If the command sequence is extended to
"SENSe:LIST:POWer:SET ON,ON,ON,IMM,POS,0,0"
"SENSe:LIST:POWer?
935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,10dB,NORM,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,20dB,NORM,30kHz,100kHz,434us,0"
the result list is extended to 3 results per frequency point (peak, RMS and average):
-28.3, -29.6, 1.5, -30.6, -31.9, 0.9, -38.1, -40.0, 2.3
Содержание FSMR
Страница 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Страница 4: ......
Страница 22: ......
Страница 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Страница 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Страница 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Страница 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Страница 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Страница 66: ......
Страница 86: ......
Страница 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Страница 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Страница 144: ......
Страница 150: ......
Страница 156: ......
Страница 202: ......
Страница 208: ......
Страница 216: ......
Страница 230: ......
Страница 254: ......
Страница 280: ......
Страница 348: ......
Страница 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Страница 390: ......
Страница 408: ......
Страница 428: ......
Страница 446: ......
Страница 536: ......
Страница 799: ......
Страница 921: ......
Страница 935: ......
Страница 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...