SENSe:SWEep Subsystem
R&S FSMR
1313.9723.12
6.1-194
E-1
This command determines the polarity of the external gate signal. The setting is valid both for the
edge of an edge-triggered signal and the level of a level-triggered signal.
This command is not available for GSM/EDGE measurements modulation accuracy
(MAC), phase/frequency error (PFE), power versus time (PVT ), carrier power (CPW),
spectrum due to switching (TRA) and spurious emissions (SPU).
Example:
"SWE:EGAT:POL POS"
Characteristics:
*RST value:
POSitive
SCPI:
device-specific
[SENSe<1|2>:]SWEep:EGATe:SOURce
EXTernal | IFPower
This command toggles between external gate signal and IF power signal as a signal source for the
gate mode. If an IF power signal is used, the gate is opened as soon as a signal at > -20 dBm is
detected within the IF path bandwidth (50 MHz).
Selection RFPower is only available with option FSP-B6 (TV and RF Trigger).
This command is not available for GSM/EDGE measurements modulation accuracy
(MAC), phase/frequency error (PFE), power versus time (PVT ), carrier power (CPW),
spectrum due to switching (TRA) and spurious emissions (SPU).
Example:
"SWE:EGAT:SOUR IFP"
'Switches the gate source to IF power.
Characteristics:
*RST value:
IFPower
SCPI:
device-specific
[SENSe<1|2>:]SWEep:EGATe:TYPE
LEVel | EDGE
This command sets the type of triggering (level or edge) by the external gate signal.
The gate opening time cannot be defined with the parameter
EGATe:LENGth
in case of level
triggering. The gate is closed when the gate signal disappears.
This command is not available for GSM/EDGE measurements modulation accuracy
(MAC), phase/frequency error (PFE), power versus time (PVT ), carrier power (CPW),
spectrum due to switching (TRA) and spurious emissions (SPU).
Example:
"SWE:EGAT:TYPE EDGE"
Characteristics:
*RST value:
EDGE
SCPI:
device-specific
[SENSe<1|2>:]SWEep:IF:SHIFt:FREQuency
<numeric value>
This command defines the signal frequency to automatically select the suitable 1
st IF
shifting for mode
IF SHIFT AUTO (SENS:SWE:IF:SHIFT:MODE AUTO).
The 1
st
IF shifting is automatically done for ACP measurements if center frequency ( =
signal frequency) is in the range of 2270 MHz … 2350 MHz. The IF SHIFT setting is
therefore ignored for ACP measurements.
Example:
"SENS:SWE:IF:MODE AUTO"
'
activates IF SHIFT AUTO
"SENS:SWE:IF:SHIF:FREQ 2GHz
'
carrier signal at 2 GHz
"
Characteristics:
*RST value:
1 GHz
SCPI:
device-specific
Содержание FSMR
Страница 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Страница 4: ......
Страница 22: ......
Страница 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Страница 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Страница 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Страница 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Страница 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Страница 66: ......
Страница 86: ......
Страница 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Страница 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Страница 144: ......
Страница 150: ......
Страница 156: ......
Страница 202: ......
Страница 208: ......
Страница 216: ......
Страница 230: ......
Страница 254: ......
Страница 280: ......
Страница 348: ......
Страница 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Страница 390: ......
Страница 408: ......
Страница 428: ......
Страница 446: ......
Страница 536: ......
Страница 799: ......
Страница 921: ......
Страница 935: ......
Страница 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...