R&S FSMR
CALCulate:MARKer Subsystem
1313.9723.12
6.1-31
E-1
CALCulate<1|2>:LIMit<1 to 8>:ACPower:ALTernate<1 to 11>:RESult?
This command queries the result of the limit check for the alternate adjacent channels in the selected
measurement window for adjacent channel power measurements.
The numeric suffix after
ALTernate<1 to 11>
denotes the alternate channel. The numeric
suffixes <1 to 8> are irrelevant for this command.
If the power measurement of the adjacent channel is switched off, the command produces a query
error.
Parameter:
The result is returned in the form <result>, <result> where
<result> = PASSED | FAILED and where the first (second) returned value
denotes the lower (upper) alternate adjacent channel.
Example:
"
CALC:LIM:ACP:ALT2 30DB, 30DB"
'Sets the relative limit value in screen A for the
'power in the lower and upper second alternate
'adjacent channel to 30 dB below the channel
'power.
"
CALC:LIM:ACP:ALT2:ABS -35DBM, -35DBM"
'Sets the absolute limit value in screen A for the
'power in the lower and upper second alternate
'adjacent channel to -35 dBm.
"
CALC:LIM:ACP ON"
'Switches on globally the limit check for the
'channel/adjacent channel measurement in
'screen A.
"
CALC:LIM:ACP:ALT:STAT ON"
'Switches on the limit check for the adjacent
'channels in screen A.
"INIT;*WAI"
'Starts a new measurement and waits for the
'sweep end.
"
CALC:LIM:ACP:ALT:RES?"
'Queries the limit check result in the second
'alternate adjacent channels in screen A.
Characteristics:
*RST value:
--
SCPI:
device-specific
This command is a query and therefore has no *RST value.
CALCulate<1|2>:LIMit<1 to 8>:ACPower[:STATe]
ON | OFF
This command switches on and off the limit check for adjacent channel power measurements in the
selected measurement window. The commands
CALCulate:LIMit:ACPower:ACHannel:STATe
or
CALCulate:LIMit:ACPower:ALTernate:STATe
must be used in addition to specify whether the limit
check is to be performed for the upper/lower adjacent channel or for the alternate adjacent channels.
The numeric suffixes <1 to 8> are irrelevant for this command.
Example:
"
CALC:LIM:ACP ON"
'Switches on the ACP limit check in screen A.
Characteristics:
*RST value:
OFF
SCPI:
device-specific
Содержание FSMR
Страница 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Страница 4: ......
Страница 22: ......
Страница 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Страница 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Страница 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Страница 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Страница 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Страница 66: ......
Страница 86: ......
Страница 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Страница 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Страница 144: ......
Страница 150: ......
Страница 156: ......
Страница 202: ......
Страница 208: ......
Страница 216: ......
Страница 230: ......
Страница 254: ......
Страница 280: ......
Страница 348: ......
Страница 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Страница 390: ......
Страница 408: ......
Страница 428: ......
Страница 446: ......
Страница 536: ......
Страница 799: ......
Страница 921: ......
Страница 935: ......
Страница 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...