MKR FCTN - Analyzer
R&S FSMR
1166.3363.12
4.14-2
E-2
Activating the Markers
Menu
MKR FCTN
:
SELECT
MARKER
The
SELECT MARKER
softkey activates the numerical selection of the
marker in the data entry field. Delta marker 1 is selected by input of ' 0 '.
If the marker is switched off, then it is switched on and can be moved later
on.
IEC/IEEE-bus command:
CALC:MARK1 ON;
CALC:MARK1:X <value>;
CALC:MARK1:Y?
PEAK
The
PEAK
softkey sets the active marker/delta marker to the peak of the
trace.
IEC/IEEE-bus command:
CALC:MARK1:MAX
CALC:DELT1:MAX
Measurement of Noise Density
NOISE MEAS
The
NOISE MEAS
softkey switches the noise measurement for the active
marker on or off. The corresponding marker becomes the
NORMAL
marker.
During noise measurement, the noise power density is measured at the
position of the marker. In the time domain mode, all points of the trace are
used to determine the noise power density. When measurements are
performed in the frequency domain, two points to the right and left of the
marker are used for the measurement to obtain a stable result.
The noise power density is indicated in the marker field. With a logarithmic
amplitude units (dBm, dBmV, dBm
8
V, dB
8
A) the noise power density is
output in dBm/Hz i.e. as level in 1 Hz bandwidth with reference to 1 mW.
With linear amplitude units (V, A, W) the noise voltage density is evaluated
in
8
V/ Hz, the noise current density in
8
A/ Hz or the noise power density
in
8
W/Hz.
The following settings have to be made to ensure that the power density
measurement yields correct values:
Detector:
Sample or RMS
Video bandwidth:
0.1 x resolution bandwidth with sample detector
(corresponds to RBW / VBW NOISE)
3 x resolution bandwidth with RMS detector
(corresponds to RBW / VBW SINE)
In the default setting, the R&S FSMR uses the sample detector for the
noise function.
With the sample detector, the trace can additionally be set to AVERAGE to
stabilize the measured values. With RMS detector used, trace averaging
must not be used since in this case it produces too low noise levels which
cannot be corrected. Instead, the sweep time can be increased to obtain
stable measurement results.
Содержание FSMR
Страница 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Страница 4: ......
Страница 22: ......
Страница 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Страница 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Страница 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Страница 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Страница 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Страница 66: ......
Страница 86: ......
Страница 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Страница 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Страница 144: ......
Страница 150: ......
Страница 156: ......
Страница 202: ......
Страница 208: ......
Страница 216: ......
Страница 230: ......
Страница 254: ......
Страница 280: ......
Страница 348: ......
Страница 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Страница 390: ......
Страница 408: ......
Страница 428: ......
Страница 446: ......
Страница 536: ......
Страница 799: ......
Страница 921: ......
Страница 935: ......
Страница 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...