R&S FSMR
Measurement Receiver Mode
1166.3363.12
4.4-12
E-2
MEAS TIME/
AVERAGE
The
MEAS TIME/ AVERAGE
softkey opens a list in which the measurement
time can be set. Results become more stable with extended measurement time,
particularly if signals with low power are measured.
MEAS TIME
SHORT (AUTO AVG, NOISE 0.05 dB, MAX TIME 1 s)
NORMAL (AUTO AVG, NOISE 0.01 dB, MAX TIME 3 s)
LONG (AUTO AVG, NOISE 0.001 dB, MAX TIME 10 s)
MANUAL (MAN AVG, MEAS TIME 100 ms)
Stationary signals with high power (>-0 dBm) require only a short measurement
time to yield stable and exact results. In this case, the
SHORT
setting is
recommended because it provides the highest measurement repetition rates.
The
NORMAL
setting increases the stability of the displayed results for the
measurement of signals with low power or of modulated signals.
The
LONG
setting is recommended for signals at the lower end of the
measurement range (<-20 dBm). This setting can be used to minimize the
influence of noise.
IEC/IEEE bus command:
SENS:PMET:MTIM LONG | NORM | SHOR
NUMBER OF
READINGS
The
NUMBER OF READINGS
softkey activates the averaging of the power
reading. It opens a window for the entry of the number of readings (averagings)
to be performed after a single sweep has been started.
The softkey is only available if the MEAS TIME / AVERAGE mode is set to
manual average.
The range for the average count is 0 to 256 in binary steps (1 / 2 / 4 / 8 …). For
average count = 0 or 1, one reading is performed.
The averaging and sweep count of the trace menu are not connected.
Results become more stable with extended average, particularly if signals with
low power are measured. This setting can be used to minimize the influence of
noise in the power meter measurement.
IEC/IEEE bus command:
SENS:PMET:MTIM:AVER:STAT ON
SENS:PMET:MTIM:AVER:COUNT
<numeric_value>
DISPLAY
LOG LIN
The
DISPLAY LOG/LIN
softkey switches the result display between logarithmic
display (units: dBm and dB) and linear display (units: W and %).
IEC/IEEE bus command:
UNIT:PMET:POW DBM
Fig. 4.4-8
Powersensor with relative display
Содержание FSMR
Страница 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Страница 4: ......
Страница 22: ......
Страница 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Страница 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Страница 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Страница 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Страница 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Страница 66: ......
Страница 86: ......
Страница 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Страница 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Страница 144: ......
Страница 150: ......
Страница 156: ......
Страница 202: ......
Страница 208: ......
Страница 216: ......
Страница 230: ......
Страница 254: ......
Страница 280: ......
Страница 348: ......
Страница 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Страница 390: ......
Страница 408: ......
Страница 428: ......
Страница 446: ......
Страница 536: ......
Страница 799: ......
Страница 921: ......
Страница 935: ......
Страница 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...