MEAS - Analyzer
R&S FSMR
1166.3363.12
4.16-4
E-2
MEAN
The
MEAN
softkey switches on the calculation of the mean value from the points
of the displayed trace or a segment of it. The linear mean value of the equivalent
voltages is calculated.
This can be used for instance to measure the mean power during a GSM burst.
For the maximum peak, the largest mean value obtained since the activation of
MAX HOLD ON
is displayed.
With
AVERAGE ON
, the mean values of a trace are averaged over several
sweeps and displayed.
The number of sweeps over which the average or the maximum value is
calculated is set with the
NUMBER OF SWEEPS
softkey.
IEC/IEEE-bus command:
CALC:MARK:FUNC:SUMM:MEAN ON
CALC:MARK:FUNC:SUMM:MEAN:RES?
STANDARD
DEVIATION
The
STANDARD DEVIATION
softkey switches on the calculation of the standard
deviation of trace points from the mean value and outputs them as measured
value. The measurement of the mean power is automatically switched on at the
same time.
For the maximum peak, the largest standard deviation obtained since the
activation of
MAX HOLD ON
is displayed.
With
AVERAGE ON
, the standard deviations of a trace are averaged over several
sweeps and displayed.
The number of sweeps over which the average or the maximum value is
calculated is set with the
NUMBER OF SWEEPS
softkey.
IEC/IEEE-bus command:
CALC:MARK:FUNC:SUMM:SDEV ON
CALC:MARK:FUNC:SUMM:SDEV:RES?
LIMIT
ON OFF
The
LIMIT ON/OFF
softkey selects the limited (
ON
) or non-limited (
OFF
)
evaluation range.
The evaluation range is defined by the
START LIMIT
and
STOP LIMIT
softkeys. If
LIMIT = ON
, signals are only searched between the two lines.
If only one limit line is switched on, time line 1 is the lower limit and the upper limit
corresponds to the stop frequency. If time line 2 is also switched on, it defines the
upper limit.
If no limit line is switched on, the evaluation range is not limited.
The default setting is
LIMIT = OFF
.
IEC/IEEE-bus command:
CALC:MARK:X:SLIM OFF
START
LIMIT
The
START LIMIT
softkey activates the entry of the lower limit of the evaluation
range.
IEC/IEEE-bus command:
CALC:MARK:X:SLIM:LEFT <value>
Содержание FSMR
Страница 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Страница 4: ......
Страница 22: ......
Страница 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Страница 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Страница 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Страница 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Страница 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Страница 66: ......
Страница 86: ......
Страница 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Страница 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Страница 144: ......
Страница 150: ......
Страница 156: ......
Страница 202: ......
Страница 208: ......
Страница 216: ......
Страница 230: ......
Страница 254: ......
Страница 280: ......
Страница 348: ......
Страница 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Страница 390: ......
Страница 408: ......
Страница 428: ......
Страница 446: ......
Страница 536: ......
Страница 799: ......
Страница 921: ......
Страница 935: ......
Страница 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...