MKR – Analyzer
R&S FSMR
1166.3363.12
4.13-4
E-1
The
REFERENCE FIXED
softkey defines the level and the fre-
quency or time of
MARKER 1
as a reference for one or several
delta markers. The measured values for one or several
markers displayed in the marker info field are derived from this ref-
erence point instead of the current values of the reference marker
(
MARKER
1).
On actuating the softkey, reference fixed is switched on and thus,
the level value and the frequency, time or x-level value of
MARKER
1
immediately become the reference point.
Additionally, the
REFERENCE
FIXED
softkey opens the submenu
where it is possible to determine manually a reference point with
level and frequency, time or x-axis level, to define a level offset or
deactivate the reference point.
The
REFERENCE FIXED
function is useful for the measurement of
the harmonic suppression at small span (fundamental not repre-
sented).
REF FIXED
ON OFF
The
REF FXD ON/OFF
softkey switches on or off the relative
measurement to a fixed reference value (
REFERENCE POINT
)
independent of the trace.
IEC/IEEE-bus command:
CALC:DELT2:FUNC:FIX ON
REF POINT
LEVEL
The
REF POINT
LEVEL
softkey enters a reference level independ-
ent of the reference marker level. All relative level values of the
delta markers refer to this reference level.
IEC/IEEE-bus command:
CALC:DELT2:FUNC:FIX:RPO:Y -10dBm
REF POINT
LVL OFFSET
The
REF POINT LVL OFFSET
softkey specifies a level offset rele-
vant to the reference level. The relative level values of the delta
markers refer to the reference point level plus the level offset.
The level offset is set to 0 dB on enabling the
REFERENCE FIXED
or
PHASE NOISE
function.
IEC/IEEE-bus command:
CALC:DELT2:FUNC:FIX:RPO:Y:OFFS 0dB
REF POINT
FREQUENCY
With the
REF POINT FREQUENCY
softkey a reference frequency
can be manually activated for the delta markers when the
REFERENCE FIXED
or
PHASE NOISE
function is used.
IEC/IEEE-bus command:
CALC:DELT2:FUNC:FIX:RPO:X 10.7MHz
Содержание FSMR
Страница 1: ...R S FSMR Measuring Receiver Operating Manual Test Measurement Operating Manual 1313 9723 12 01...
Страница 4: ......
Страница 22: ......
Страница 26: ...Contents Preparing for Operation R S FSMR 1155 5047 12 I 1 2 E 2 Fig 1 1 Front View...
Страница 28: ...Front View R S FSMR 1166 3363 12 1 2 E 2 Fig 1 1 Front View...
Страница 30: ...Front View R S FSMR 1166 3363 12 1 4 E 2 Fig 1 1 Front View...
Страница 32: ...Front View R S FSMR 1166 3363 12 1 6 E 2 Fig 1 1 Front View...
Страница 34: ...Front View R S FSMR 1166 3363 12 1 8 E 2 Fig 1 1 Front View...
Страница 66: ......
Страница 86: ......
Страница 130: ...Menu Overview R S FSMR 1166 3363 12 3 42 E 3 Menu Overview another Keys MKR Key...
Страница 141: ...R S FSMR Menu Overview 1166 3363 12 3 53 E 3 LOCAL Menu LOCAL...
Страница 144: ......
Страница 150: ......
Страница 156: ......
Страница 202: ......
Страница 208: ......
Страница 216: ......
Страница 230: ......
Страница 254: ......
Страница 280: ......
Страница 348: ......
Страница 376: ...Configuration Setup R S FSMR 1166 3363 12 4 19 28 E 2...
Страница 390: ......
Страница 408: ......
Страница 428: ......
Страница 446: ......
Страница 536: ......
Страница 799: ......
Страница 921: ......
Страница 935: ......
Страница 937: ...Contents Error Messages R S FSMR 1166 3363 12 I 9 2 E 1...