Remote Control Commands
R&S
®
NRX
298
User Manual 1178.5566.02 ─ 07
<resolution_or_source_list>
<source_list>
<expr>
Usage:
Query only
CONFigure<Measurement>:ARRay[:POWer][:AVG]:RATio?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
FETCh<Measurement>:ARRay[:POWer][:AVG]:RATio?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
READ<Measurement>:ARRay[:POWer][:AVG]:RATio?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
MEASure<Measurement>:ARRay[:POWer][:AVG]:RATio?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
CALCulate<Measurement>:MATH[:EXPRession]
"(SENSn /SENSm)"
.
Ratio measured by two power sensors with buffering. The used parameters are descri-
bed in
Suffix:
<Measurement>
.
1 to 8
Measurement
Query parameters:
<buffered_size>
<expr>
<expected_value_or_source_list>
<resolution_or_source_list>
<source_list>
<expr>
Usage:
Query only
CONFigure<Measurement>:ARRay[:POWer][:AVG]:RATio:RELative?
<buffered_size>[, <expected_value_or_source_list>,
<resolution_or_source_list>, <source_list>...]
FETCh<Measurement>:ARRay[:POWer][:AVG]:RATio:RELative?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
READ<Measurement>:ARRay[:POWer][:AVG]:RATio:RELative?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
MEASure<Measurement>:ARRay[:POWer][:AVG]:RATio:RELative?
<buffered_size>[, <expected_value_or_source_list>,
<resolution_or_source_list>, <source_list>...]
CALCulate<Measurement>:MATH[:EXPRession]
.
Relative ratio measured by two power sensor with buffering. The used parameters are
described in
Suffix:
<Measurement>
.
1 to 8
Measurement
Calculation Functions
www.allice.de
Allice Messtechnik GmbH