Remote Control Commands
R&S
®
NRX
354
User Manual 1178.5566.02 ─ 07
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<state>
*RST:
OFF
CALCulate<Measurement>[:CHANnel<Channel>]:ROSCillator:SOURce
<source>
Effective for R&S
NRP series power sensors and R&S
frequency selective power sen-
sors.
Sets the source of the reference oscillator.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<source>
HOST | INTernal | REFio
*RST:
INTernal
[SENSe<Sensor>:]BANDwidth[:RESolution]:TYPE:AUTO[:STATe]
<state>
Effective for R&S
frequency selective power sensors.
If enabled, sets the filter type suitable for the currently chosen measurement mode and
bandwidth.
You can query the selected filter type using
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
*RST:
ON
[SENSe<Sensor>:]BANDwidth[:RESolution]:TYPE[:VALue]
<value>
Effective for R&S
frequency selective power sensors.
Sets the filter type for resolution bandwidth filter. The filter bandwidth is not affected.
If you want to set the filter type automatically, use
Configuring Sensors