Remote Control Commands
R&S
®
NRX
434
User Manual 1178.5566.02 ─ 07
CALCulate<Measurement>[:POWer]:TSLot[:AVG][:EXCLude]:MID[:STATe]
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
*RST:
OFF
[SENSe<Sensor>:]POWer:CCDFunction:REFerence
<ref>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:CCDF:THReshold
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<ref>
Range:
-290.0 to +110.0
*RST:
+0.0
Default unit: dBm
[SENSe<Sensor>:]POWer:PEP:HOLD
<time>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:PEP:HOLD:TIME
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
1.0e-3 to 1.0e-1
*RST:
6.0e-2
Default unit: s
[SENSe<Sensor>:]RANGe:AUTO
<state>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:RANGe:
AUTO
R&S NRP2 Compatibility
www.allice.de