Remote Control Commands
R&S
®
NRX
358
User Manual 1178.5566.02 ─ 07
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<direction>
AUTO | FORWard | REVerse
AUTO
Determines the power flow direction automatically. The greater
value of two measured values is automatically assigned as for-
ward power.
FORWard | REVerse
Sets a fixed direction of the forward power, either from port 1 to
port 2 (
FORWard
), or from port 2 to port 1(
REVerse
).
*RST:
AUTO
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:PEP:HOLD:TIME
<value>
Sets the hold time of the peak hold circuit of the power sensor.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
1e-3 to 1e-1
*RST:
0.01
Default unit: s
Manual operation:
See
INPut<Sensor>:PORT:SOURce:AUTO
<auto>
Enables or disables the automatic assignment of the forward direction.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<auto>
ON
The direction in which the greater power flows is taken as the
forward direction.
OFF
The forward direction is defined by
Configuring Sensors