Remote Control Commands
R&S
®
NRX
359
User Manual 1178.5566.02 ─ 07
*RST:
0
Manual operation:
See
INPut<Sensor>:PORT:SOURce[:VALue]
<val>
INPut<Sensor>:PORT:SOURce:AUTO
OFF
is set.
Sets a fixed direction of the forward power.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<val>
1
Port
1 to port
2
2
Port
2 to port
1
Range:
1 to 2
*RST:
1
Manual operation:
See
13.7.8.2
NRT Correction Settings
Further information:
●
Chapter 8.5.2, "NRT Correction Settings"
..........................................................................................359
.................................................................................... 360
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet:RPLane
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:DMODulation[:VALue]
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:DMODulation:WCDMa:CRATe
........................................................................................ 361
.......................................................................................361
[SENSe<Sensor>:]DM:STATe
<state>
Requires the sensor interface for R&S NRT (R&S NRX-B9).
Enables or disables the modulation correction. Set the communication standard using
.
Suffix:
<Sensor>
.
101
R&S NRT directional power sensor
Parameters:
<state>
OFF | ON
*RST:
0
Configuring Sensors
www.allice.de
Allice Messtechnik GmbH