Getting Started
R&S
®
NRX
22
User Manual 1178.5566.02 ─ 07
3.1.8.2
Optional Sensor Interface for R&S NRT (R&S NRX-B9)
See
Suitable for R&S
NRT directional power sensors.
Communication between R&S NRT-Zxx power sensor and R&S NRX is only possible
with a baud rate setting of 38400
Bd. This setting is the factory default that must be
restored if the setting was changed. If the R&S NRT-Zxx power sensor is not recog-
nized by the R&S NRX, check that the baud rate setting of the R&S NRT-Zxx power
sensor is 38400
Bd. See the manual of the R&S NRT-Zxx power sensor for details.
1. Connect the R&S NRT-Zxx power sensor between source and load.
a) Connector (1) to the source.
b) Connector (2) to the load.
2. Connect the cable of the R&S NRT-Zxx power sensor (3) to the sensor interface for
R&S NRT (R&S NRX-B9).
3.1.8.3
LAN Interface
See
Chapter 3.2.2.2, "Ethernet Interface"
Suitable for LAN power sensors.
R&S power sensors that are connected to the LAN interface are not recognized auto-
matically. Add them, see
Preparing for Use