Remote Control Commands
R&S
®
NRX
363
User Manual 1178.5566.02 ─ 07
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt[:VALue]
<value>
Effective if
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:
OFF
is set.
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Range:
1 to 1048576
*RST:
4
Manual operation:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:ENUM
<value>
See
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt[:
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
E1 | E2 | E4 | E8 | E16 | E32 | E64 | E128 | E256
*RST:
E4
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:APERture:MODE
<mode>
Specifies which integration time is used for a single measurement.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<mode>
DEFault | USER
Configuring Sensors