Remote Control Commands
R&S
®
NRX
404
User Manual 1178.5566.02 ─ 07
Example:
SYSTem:ERRor:CODE
Queries the oldest entry in the error queue.
Response: 0
No errors have occurred since the error queue was last read out.
Usage:
Query only
SYSTem:ERRor:COUNt?
Queries the number of entries in the error queue.
Return values:
<ErrorCount>
Example:
SYSTem:ERRor:COUNt
Queries the number of entries in the error queue.
Response: 1
One error has occurred since the error queue was last read out.
Usage:
Query only
SYSTem:ERRor:EXTended[:STATe]
<state>
Enables or disables the extended error messages that provide more information than
. The location of errors in the parsed command strings is
also shown.
Parameters:
<state>
*RST:
ON
SYSTem:ERRor[:NEXT]?
Queries the error/event queue for the oldest item and removes it from the queue. The
response consists of an error number and a short error description.
Return values:
<ErrorCode>
Positive error numbers are instrument-dependent. Negative
error numbers are reserved by the SCPI standard.
0
No errors have occurred since the error queue was last read out.
<ErrorDescription>
Short error description
Usage:
Query only
SYSTem:SERRor[:ALL]?
Queries all errors from the list of static errors.
Usage:
Query only
System Information and Configuration
www.allice.de
Allice Messtechnik GmbH