Remote Control Commands
R&S
®
NRX
375
User Manual 1178.5566.02 ─ 07
OUTPut:RECorder<output>:LIMit:UPPer:RATio:SWR
<value>
Effective if
REC
,
FREC
or
RREC
Sets the standing wave ratio (SWR) measurement value that corresponds to 2.5
V out-
put voltage.
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-1e18 to 1e18
*RST:
10.0
Default unit: -
Manual operation:
See
OUTPut:RECorder<output>:LIMit:UPPer:RATio[:VALue]
<value>
Effective if
REC
,
FREC
or
RREC
Sets the power ratio measurement value that corresponds to 2.5
V output voltage.
If you enter a value without unit, the unit is defined by
. For further information, see
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-180.0 to +180.0
*RST:
+10.0
Default unit: dB
Manual operation:
See
OUTPut:TRIGger:SOURce
<source>
Effective if
TOUT
is set.
Sets the trigger source.
Parameters:
<source>
SENS1 | SENS2 | SENS3 | SENS4 | EXTernal | CHKSource
*RST:
EXTernal
Manual operation:
See
"Trigger Source for Trigger Output"
Configuring the Analog Signal Ouput and the Trigger Input/Output
www.allice.de
Allice Messtechnik GmbH