Sensor Configuration
R&S
®
NRX
122
User Manual 1178.5566.02 ─ 07
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt[:VALue]
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage:COUNt[:
VALue]
Recalc Filter Length
Available for continuous average, burst average, time gate measurements.
Recalculates the number of readings that are averaged for one measured value.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO[:
STATe]
Clear Filter Buffer
Available for continuous average, burst average, time gate measurements.
Clears the filter buffer.
Remote command:
Fixed Noise Mode
Available for continuous average, burst average measurements.
Sets the autofilter.
"Normal"
Sets the averaging number so that the intrinsic noise of the power
sensor, 2 standard deviations, does not exceed the specified
"Fixed Noise"
Limits the averaging number as specified in
to avoid very long settling times.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:
TYPE
Noise Content
is set to "Normal", available for continuous average, burst average
measurements.
Sets the averaging number so that the intrinsic noise of the power sensor does not
exceed the specified value.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:
NSRatio
Maximum Settling Time
is set to "Fixed Noise", available for continuous average, burst
average measurements.
Sets an upper time limit, a maximum time, that is never exceeded.