Remote Control Commands
R&S
®
NRX
338
User Manual 1178.5566.02 ─ 07
CALCulate<Measurement>[:CHANnel<Channel>]:INPut:ATTenuation[:VALue]
<value>
Requires an R&S
frequency selective power sensor.
Effective if
CALCulate<Measurement>[:CHANnel<Channel>]:INPut:
OFF
is set.
Sets the input attenuation.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Only two values are possible, 0.0
dB and 30.0
dB. The entered
value is rounded to the next value.
Range:
0.0 to 30.0
*RST:
30.0
Default unit: dB
Manual operation:
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:RANGe:AUTO
<value>
Enables or disables the automatic measurement path selection.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
*RST:
ON
Manual operation:
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:RANGe:CLEVel:
STATe
<value>
Effective if
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:
ON
is set.
Enables or disables the reduction of the transition range between the measurement
paths, set by
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:
Configuring Sensors