Remote Control Commands
R&S
®
NRX
437
User Manual 1178.5566.02 ─ 07
[SENSe<Sensor>:]SGAMma:CORRection:STATe
<state>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma:CORRection:
STATe
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
ON | OFF
*RST:
OFF
[SENSe<Sensor>:]SGAMma:PHASe
<phase>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma:PHASe
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<phase>
Range:
-360.0 to 360.0
*RST:
0.0
Default unit: degree
[SENSe<Sensor>:]SGAMma[:MAGNitude]
<magnitude>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma[:MAGNitude]
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<magnitude>
Range:
0.0 to 1.0
*RST:
0.0
[SENSe<Sensor>:]STATistics:SAMPles[:MINimum]
<value>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>:STATistics:SAMPles[:MINimum]
R&S NRP2 Compatibility
www.allice.de
Allice Messtechnik GmbH