Remote Control Commands
R&S
®
NRX
270
User Manual 1178.5566.02 ─ 07
CALCulate<Measurement>:STATistics:APERture
<value>
Effective for statistics measurements.
Sets the width of the sampling window for unsynchronized acquisition, that is if
CALCulate<Measurement>:STATistics:TGATe:SELection
0
is set.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
10e-6 to 10.0
*RST:
0.01
Default unit: s
Manual operation:
See
CALCulate<Measurement>:STATistics:AWGN[:STATe]
<value>
Effective for statistics measurements.
Enables or disables the internal, additional white Gaussian noise (AWGN) source. If
enabled, you cannot measure with a second power sensor.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
*RST:
OFF
Manual operation:
See
CALCulate<Measurement>:STATistics:SAMPles[:MINimum]
<value>
Effective for statistics measurements.
Sets the minimum number of samples.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
1 to 2147483647
*RST:
1000000
Manual operation:
See
CALCulate<Measurement>:STATistics:TGATe:SELection
<value>
Effective for statistics measurements.
Configures the sampling window.