Remote Control Commands
R&S
®
NRX
264
User Manual 1178.5566.02 ─ 07
Parameters:
<value>
Range:
0.0 to 15.0
*RST:
0.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG]:TIME
<value>
Sets the start time of the gate.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Gate>
1 to 4
Time gate
Parameters:
<value>
Range:
50.0e-9 to 0.1
*RST:
1.0e-3
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID:
OFFSet[:TIME]
<value>
Sets length of the fence.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Gate>
1 to 4
Time gate
Parameters:
<value>
Range:
0.0 to 0.1
*RST:
0.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID:TIME
<value>
Sets the start time of the fence. The start time refers to the start of the gate.
Suffix:
<Measurement>
.
1 to 8
Measurement
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH