Remote Control Commands
R&S
®
NRX
184
User Manual 1178.5566.02 ─ 07
ABORt<Measurement>
Immediately interrupts the current measurement. If the measurement has been started
as a single measurement (
INITiate<Measurement>[:IMMediate]
), the power
sensor goes into the idle state. However, if a continuous measurement is in progress
(
INITiate<Measurement>:CONTinuous
ON
), the trigger system of the power sen-
sor enters the waiting for trigger state, and if the trigger condition is met, a new mea-
surement is immediately started.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Event
ABORt<undef>:ALL
Applies to all connected power sensors. See
Suffix:
<undef>
.
1 to n
No suffix required.
Usage:
Event
INITiate<Measurement>:CONTinuous
<state>
Enables or disables the continuous measurement mode. In continuous measurement
mode, the power sensor does not go into the idle state after a measurement has been
completed, but immediately executes another measurement cycle.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<state>
instead of
ON
Measurements are performed continuously. If a measurement is
completed, the power sensor does not return to the idle state but
enters the waiting for trigger state again.
Starting and Ending a Measurement
www.allice.de
Allice Messtechnik GmbH