INITiate<Measurement>[:IMMediate]
Starts a single measurement cycle. The power sensor changes from the idle state to
the waiting for trigger state. As soon as the trigger condition is fulfilled, the sensor
begins the measurement. Depending on the number of trigger events that are required,
e.g. for averaging, the power sensor enters the waiting for trigger state several times.
Once the entire measurement is completed, a measurement result is available, and the
power sensor enters the idle state again.
Use the command only after the continuous measurement mode has been disabled
using
INITiate<Measurement>:CONTinuous
OFF
.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Example:
If you work in a master/slave setup, you need to trigger the slave
before triggering the master. To prevent overlapping execution,
use
*WAI
Chapter 14.3, "Command Sequence and
on page 460. In this example, sensor
2 is the
slave, sensor
1 is the master:
INIT2
*WAI
INIT1
Usage:
Event
INITiate<Undef>:ALL[:IMMediate]
Applies to all connected power sensors. See
Suffix:
<Undef>
.
1 to n
No suffix required.
Usage:
Event
13.5
Measurement Settings and Results
The R&S NRX supports up to 4
measurement channels. The trace, pulse analysis, sta-
tistics measurements support 2
traces. Thus, each measurement channel has 2
sepa-
rate measurements, adding up to 8
different measurement results that are configured
using the commands of the
CALCulate<Measurement>
subsystem.