Remote Control Commands
R&S
®
NRX
395
User Manual 1178.5566.02 ─ 07
You can query the defined string using
.
Parameters:
<string>
Identification string. Maximum string length is 128 characters.
Manual operation:
See
SYSTem:IDN:AUTO
<status>
Enables or disables the automatic instrument identification for
Parameters:
<status>
ON | OFF
*RST:
1
Manual operation:
See
SYSTem:IDN:MODE
<mode>
Sets which identification string is used.
Parameters:
<mode>
AUTO | USER
AUTO
Automatic instrument identification.
USER
Customized identification string. Define the string using
.
*RST:
AUTO
Manual operation:
See
SYSTem:LANGuage
<language>
Sets the native remote command set or an emulation of a predecessor.
Setting parameters:
<language>
String
SCPI
Native remote command set of the R&S NRX.
NRP2
Emulation of the R&S NRP2.
NRP
Emulation of the R&S NRP.
Query parameters:
<language>
String
SCPI | NRP2 | NRP
Manual operation:
See
System Information and Configuration