Remote Control Commands
R&S
®
NRX
388
User Manual 1178.5566.02 ─ 07
13.13.1
Presetting
...................................................................................388
SYSTem:PRESet
Sets the R&S NRX to a defined initial state. The default settings are indicated in the
description of commands as *RST value.
, this command corresponds to
Table 13-16: Differences between *RST and SYSTem:PRESet
Command
*RST
SYSTem:PRESet
[SENSe<Sensor>:]AVERage:TCONtrol
REPeat
MOVing
CALCulate<Measurement>[:
CHANnel<Channel>]:AVERage:TCONtrol[:ENUM]
REPeat
MOVing
CALCulate<Measurement>[:
CHANnel<Channel>]:TRACe:AVERage:
TCONtrol[:ENUM]
REPeat
MOVing
INITiate<Measurement>:CONTinuous
OFF
ON
Usage:
Event
Manual operation:
See
SYSTem:SENSor<Sensor>:RESet
Sets the selected power sensor to a defined initial state.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Event
13.13.2
Shutdown and Reboot
.................................................................................389
SYSTem:REBoot
Reboots the R&S NRX.
Usage:
Event